Purwaningsih, Nur Vita and Maulidiyanti, Ellies Tunjung Sari and Widyastuti, Rahma Comparison of serum stability against stability uric acid and glucose. Research journal of pharmacy and technology.
Comparison of Serum Stability_RJPT_VT ET RW (1).pdf
Download (384kB) | Preview
Hasil Plagiasi Comparison of Serum Stability Against Uric Acid and Glucose.pdf
Download (1MB) | Preview
Abstract
The selection of tools, especially the use of vacutainer tubes, affects the results of the examination to determine
the stability of serum from plain vacutainer with vacutainer separation gel on the quality of uric acid and glucose
test results. This study used an experimental design. The study took blood samples of 3 ml each in SST vacutainer
and plain vacutainer serum. The mean uric acid levels using plain vacutainer and SST vacutainer were 4.5875 ±
0.59048 and 4.9250 ± 0.56318, respectively. The mean glucose levels using plain vacutainer and SST vacutainer
were 90.64 ± 12.33 and 94.12 ± 13.30, respectively. Uric acid and glucose levels using gel-separated vacutainer
were higher than uric acid levels using plain vacutainer. There was a significant difference in uric acid levels
between regular vacutainers and SST vacutainers, similar to glucose levels (p = 0.000, p = 0.000). Uric acid and
glucose levels using SST vacutainer were higher than uric acid levels using plain vacutainer. The following types
of vacutainer separator gel and plain vacutainer affect the quality of uric acid and glucose test results.
KEYWORDS: Uric Acid Levels, Glucose levels, Pre-Analytics, Serum, Vacutainer Separator Gel, Vacutainer
Item Type: | Article |
---|---|
Subjects: | R Medicine > R Medicine (General) |
Divisions: | Jurnal > Fakultas Ilmu Kesehatan |
Depositing User: | RAHMA WIDYASTUTI WIDYASTUTI |
Date Deposited: | 01 Aug 2023 12:19 |
Last Modified: | 01 Aug 2023 12:19 |
URI: | https://repository.um-surabaya.ac.id/id/eprint/7390 |